Check the Arctic version in your quote
The Arctic page on cryo.com lists these alternative supplies. The page does not say whether it describes the same revision as the earlier Arctic Performance figures below, so confirm which one is quoted.
- One listed option
- 230 V50 Hz
- The other listed option
- 110 V60 Hz
Ask which option your quote supplies and obtain its circuit requirements. Do not transfer the earlier 13 A peak label to the 110 V option.
| Purchase check | What the Arctic page lists | What still needs an answer |
|---|---|---|
| Nitrogen use | About 12 L per session, described as an average | Does that include precooling, re-cooling and other liquid use? |
| Delivery weight | 549 kg with shipping crates; 305 kg without them | Which weight and package dimensions apply to each delivery stage? |
| Depth with door | 1000 mm / 41 in | Those labels do not match. Request one dimensioned drawing before planning the room. |
The older room minimums are not the new page's body dimensions. Match the product name and revision before using either set for your site.
Earlier Arctic Performance specifications
| Spec | What the maker says | Checked | What to ask |
|---|---|---|---|
| Arctic cooling | Nitrogen heat exchange | October 6, 2026 | Do not apply the electric e-Arctic connection sheet. |
| Arctic format | Adjustable window; WBC and PBC modes | October 6, 2026 | Confirm the intended configuration and operating instructions. |
| Minimum supply voltage | 220 V | October 11, 2026 | Current delivered phase/frequency still need a drawing. |
| Maximum supply voltage | 230 V | October 11, 2026 | Not a utility-consumption figure. |
| Peak current | 13 A peak, as stated by the maker | October 11, 2026 | Not average operating current. |
| Minimum room length | 2.7 m | October 11, 2026 | Not chamber-body length. |
| Minimum room width | 2.4 m | October 11, 2026 | Not an approved site layout. |
| Minimum room height | 2.6 m | October 11, 2026 | Check access and service space too. |
| Minimum door width | 90 cm | October 11, 2026 | Check the actual delivery route. |
Our take
Start by naming the exact model in every supplier question. A name that differs by one letter can change the utility and installation questions. Use the earlier Arctic Performance requirements only after matching the offered revision. Ask for a complete liquid-use estimate and the electrical drawing for that same unit.
For an electric shortlist, request the e-Arctic or Glacier configuration sheet rather than recycling the Arctic page. A remote machinery or water component has a different space boundary from the main chamber.
e-Arctic temperature labels
The e-Arctic product page gives different lower temperature endpoints:
- Technical Specification / Temperature
- -80 to -90 C-112 to -130 F
- Special features / Rapid Cooling Time / Air temperature
- -80 to -110 C-112 to -166 F
- Benefits / perceived value
- -110 C-166 F
Fahrenheit values are converted from the published Celsius figures.
Checked October 19, 2026. Benefits also labels the first range as air temperature. The page does not explain the difference. The perceived value is not an air reading or a measurement we made. Ask for the quoted revision's air range, measurement conditions and definition of perceived temperature.
Ask for the complete electric layout
| Electric-model question | e-Arctic Performance | Glacier Performance |
|---|---|---|
| Published US-frequency option | 208 V / 60 Hz / 60 A as stated by the maker | 208 V / 60 Hz; the page says a US voltage booster is required |
| Main chamber module, L x W x H | 1800 x 1000 x 2300 mm | 2100 x 1230 x 2500 mm |
| Other hardware to place | Separate machinery | Separate machinery and water-cooling module |
| Layout issue to resolve | Operating: 2310 x 2240 mm; approximate installation: 2008 x 1740 mm. Compare all three space labels before planning the room | Supplier describes water-cooling installation as outdoors. Compare three operating spaces. |
| Energy use | Current rating is not measured kWh | 30 kW is a supplier power rating, not kWh per day |
Use the figures above to screen your options, then ask for the electrical sheet and installation drawing for the exact unit offered. Confirm the complete component layout and US supply version before planning the room.
Request a drawing that identifies chamber, machinery and any water equipment, the working and installation spaces around each, connection routes and outdoor requirements if applicable. Then check voltage, frequency, phase and service scope against that same drawing. The installation guide explains the responsibility handoff.
Supplier space labels, not equipment-body dimensions.
01 / Chamber
- Length x width
- 2,544 x 2,713 mm
- Height
- 2,700 mm
02 / Machinery
- Length x width
- 2,075 x 850 mm
- Height
- 1,673 mm
03 / Water cooling
- Length x width
- 2,300 x 1,110 mm
- Height
- 1,110 mm
One scale, original L/W order; heights listed separately. These plots do not form a room plan.
Nine dimension labels checked ; manufacturer source. The water-cooling installation is described as outdoors.
Compare the three e-Arctic space labels
The operating rectangle is larger than the supplier's approximate installation rectangle. Ask which spaces apply to the quoted configuration before planning the room.
01 / Module space
- Length x width
- 1,800 x 1,000 mm
- Module-space height
- 2,300 mm
02 / Operating space
- Length x width
- 2,310 x 2,240 mm
03 / Approximate installation space
- Length x width
- 2,008 x 1,740 mm
All three plan rectangles use the same scale and original L/W order. Height is listed separately. The outlines do not assign clearances to particular sides or combine these labels into a room.
Dimension fields checked ; supplier dimension labels. The separate machinery, connections and service access still need the configured installation drawing.
Compare on the same basis
The electric/nitrogen cost comparison, full equipment price index and manufacturer document directory cover different next steps. For another electric document set, read MECOTEC; for nitrogen supply, use the local nitrogen worksheet.